Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20688 |
| Brand Name: | Monaco RTP |
| Model: | |
| Affected: | Creating 3D plans and Motorized Wedges |
| Software version: | V5.10 or V5.20 |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | When creating 3D plans using either MU or Dose weighting modes, if the user changes the Physician's Intent Rx Dose and the number of fractions, and then modifies the wedge angle, the MU value is scaled incorrectly. The scaling of the MU is in direct proportion to the fractional change. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/10/2016 |