Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20675
Brand Name:Maquet operating table accessories
Model:
Affected:Accessory Adaptor (600525A0)
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet GmbH
Issue: Investigation has identified that the raw material of the accessory adapter does not meet specifications. The tensile and the flexural strength of the used material are less than those of the specified materials. Serious injury may occur should the attached accessory release from the adapter without warning during a surgical procedure.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/9/2016