Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20675 |
| Brand Name: | Maquet operating table accessories |
| Model: | |
| Affected: | Accessory Adaptor (600525A0) |
| Software version: | |
| Recalling Organisation: | Maquet Australia Pty Ltd (NZ) 12G Andromeda Cres East Tamaki Auckland 2013 |
| Contact Information: | 021 874 882 |
| Manufacturer: | Maquet GmbH |
| Issue: | Investigation has identified that the raw material of the accessory adapter does not meet specifications. The tensile and the flexural strength of the used material are less than those of the specified materials. Serious injury may occur should the attached accessory release from the adapter without warning during a surgical procedure. |
| Recall action type: | Recall |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/9/2016 |