Detail of Recall

Type of Product:Device
Medsafe Reference:20673
Brand Name:AutomaticQC Cartridge - Positive Bias tHb Level 1 AQC
Model:
Affected: Lots AQC/19516, AQC/19616, AQC/20816, AQC/20916, AQC/21016, AQC/21416, AQC/21516
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that the lots of AQC have been incorrectly value assigned resulting in a +8 g/L bias for Level 1 tHb on RAPIDPoint 400/405/500 systems. Some users are reporting higher than expected results for this AQC level
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/9/2016