Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20640 |
| Brand Name: | Radiometer ABL90Flex/ ABL90Flex Plus Blood Gas Analysers |
| Model: | |
| Affected: | |
| Software version: | ALL running software below ver3.2 |
| Recalling Organisation: | Radiometer Pacific Suite 1 1/30 Benedicts Street Eden Terrace Auckland 1010 |
| Contact Information: | 0800 723 722 |
| Manufacturer: | Radiometer Medical ApS |
| Issue: | Potential risk that the analyser may report patient sample results although the analyser performance has not been verified by running quality controls. This situation may occur if the customer selects the "exit conditioning" option when a new Sensor Cassette is loaded. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/9/2016 |