Detail of Recall

Type of Product:Device
Medsafe Reference:20640
Brand Name:Radiometer ABL90Flex/ ABL90Flex Plus Blood Gas Analysers
Model:
Affected:
Software version:ALL running software below ver3.2
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:Potential risk that the analyser may report patient sample results although the analyser performance has not been verified by running quality controls. This situation may occur if the customer selects the "exit conditioning" option when a new Sensor Cassette is loaded.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/9/2016