Detail of Recall

Type of Product:Device
Medsafe Reference:20635
Brand Name:Philips Brilliance 64 Ingenuity CT/Core/Core128
Model:
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:While performing High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as:
-Non uniformity of the image, -CT number shift, greater than 5 Hounsfield units,-Reduction in grey/white matter differentiation. The issue is most severe in the high resolution mode
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2016