Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20634
Brand Name:Plum 360 Infusion pump
Model:
Affected:Plum 360 infusion pump
Software version:
Recalling Organisation:Hospira NZ Ltd
58 Richard Pearse Drive
Airport Oaks
Mangere
Auckland 2022
Contact Information:04 384 7463
Manufacturer:Hospira Inc
Issue:An issue has been indentified with Plum 360 infuser where the User Interface does not update to reflect a calculated dose value. This only occurs when using a dosing unit in the mmol or units families. The design is such that when a user enters a rate or volume to be infused (VTBI) the pump calculates the dose upon entry of every digit.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/9/2016