Detail of Recall

Type of Product:Device
Medsafe Reference:20631
Brand Name:Topcon Pascal Synthesis and Pascal Synthesis TwinStar (only with EpM software enabled)
Model:
Affected:
Software version:1.0.0, 1.0.1, 2.0.0, 3.1.1, 3.2.0, 3.3.0, 3.3.1 and 3.3.2) and 1.0.0, 1.0.1, 2.0.0, 3.1.1, 3.2.0, 3.3.0, 3.3.1 and 3.3.2)
Recalling Organisation:Optimed NZ Ltd
Unit 1
3 Northside Drive
Whenuapai
Auckland 0814
Contact Information:
Manufacturer:Topcon Medical Laser Systems Inc
Issue:An anomaly has been discovered with titration, only when the EpM software feauture is turrned on during the titration process.
This event may occur if the user turns on EpM, turns off the Landmark (LM) function (with any EpM supported pattern) and then enters titration mode. In either single or multiple spot titration patterns, the power output upon pressing the footswitch is the same as the EpM power. The EpM power is a percentage of the power selected in the EpM control. The user would expect the power output during titration to be equal to 100% of the selected power level and may increase the power level during the titration process. When the expected tissue response is not observed, the power level resulting from titration will produce an over-power treatment when the user subsequently treats with the titration turned off.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/10/2016