Detail of Recall

Type of Product:Device
Medsafe Reference:20623
Brand Name:Siemens RAPIDPoint® 400 and and RAPIDPoint® 405 Measurement Cartridge
Model:0329756, 10341161, 10844812, 10283222, 10313971 and 10310469
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:
Issue:An error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge serial numbers 2610900214 through to 2621001325 has been identified. At extremely high chloride values (>122nmol/L) a positive bias of 5% or greater may be observed. Chloride values should be interpreted in the context of all serum electrolytes (sodium, bicarbonate) and acid base status.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/9/2016