Detail of Recall

Type of Product:Device
Medsafe Reference:20622
Brand Name:Siemens Dimension IRON and Dimension Vista IRON
Model:DF85 and K3085
Affected:All lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Current Instructions for Use ( IFU) information under the non-interference section related to desferrioxamine is incorrect.
In rare circumstances, where serum iron may be measured following initial administration of desferrioxamine, unexpected interference leading to substantial depression of serum iron values may ultimately lead to inadequate desferrioxamine dosing.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/9/2016