Detail of Recall

Type of Product:Device
Medsafe Reference:20610
Brand Name:RaySearch Radiation Therapy Treatment Planning System
Model:2.5, 3.0, 4.0, 4,5, 5.0 and 4.3 and InverseArc 1.0
Affected:
Software version:
Recalling Organisation:CARSL Consulting
30 Fulford Road
Havelock North 4130
Hastings
Contact Information:09 238 3447
Manufacturer:RaySearch Laboratories AB (publ)
Issue:Two issues have been identified with the dose calculation when using a region of interest (ROI) of type Fixation or Support with material override within the patient outline in RayStation models.
1. When an ROI with material override overlaps a Fixation or Support ROI, the material used for dose calculation in the overlapping region may not be as intended.
2. When a Support or Fixation ROI overlaps the patient outline (External ROI) in a dose grid voxel along the outer surface of the patient outline, the density used for dose calculation in that voxel may not be as intended.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/9/2016