Detail of Recall

Type of Product:Device
Medsafe Reference:20604
Brand Name:Propax NeuroPak Procedure Pak
Model:
Affected:2902463-04, 72658-00036-02
Software version:
Recalling Organisation:BSN Medical Ltd - T/A Essity New Zealand
103 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:09 846 5047
Manufacturer:BSN Medical Ltd - T/A Essity New Zealand
Issue:IFU being updated due to complaints from the users in relation to breaking of the instrument at the junction of the handle and the metal tubing when pressure is applied at the tip. If the issue were to occur it would result in a loss of suction and loss of visibility in the surgical field.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/9/2016