Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20582
Brand Name:Heater Cooler Unit
Model:
Affected:HU 35, HCU 20, HCU 30 & HCU 40
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet Cardio Pulmonary GmbH
Issue:Maquet Cardiopulmonary have received reports confirming the presence of Mycobacterial contamination (and other bacterial contaminants) in the system water of Heater and Heater/Cooler devices, mainly used during cardiothoracic surgeries to warm or cool a patient. Although the water in the circuits does not come into direct contact with the patient, there is the potential for contaminated water to enter other parts of the device or transmit bacteria through the air (aerosolize) through the device's exhaust vent into the environment and to the patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/8/2016