Detail of Recall

Type of Product:Device
Medsafe Reference:20579
Brand Name:Etest COLISTIN CO 256 WW B30 & B100
Model:
Affected:Multiple lots
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:The manufacturer was alerted about a potential underestimation of the MIC values for Enterobacteriaceae and Acinetobacter, a complaint investigation to confirm product performance and determine. While the internal and external investigations are still ongoing, to date and based on data available, the following have been identified.
For the Enterobacteriaceae strains, the first results of our investigation state that the performance of EtestR COLISTIN CO256 conforms if used in conjunction with the MHE aga.
There is a potential of False Susceptible results on EtestR COLISTIN CO256 WW with some Mueller Hinton agar.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/8/2016