Detail of Recall

Type of Product:Device
Medsafe Reference:20575
Brand Name:NucliSENS easyMag Magnetic Silica Mag SIL
Model:
Affected:Z017BB1MS, Z017BF1MS, Z017CC1MS, Z017BA1MS,Z017CG1MS, Z017DC1MS, Z017DD1MS
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:Following customer complaints about amplification performance issues with the home brew Polymerase Chain Reaction (PCR) and Real Time PCR assays. The investigation concluded that downstream applications were impacted when the extraction volume was higher than 400 ìL and for double stranded nucleic acids target.
Double stranded nucleic acid applications with small (< 40Kbp) and medium genome sizes (< to 1200 Kbp) i.e DNA Viruses are more impacted than higher human genomic DNA and bacterial applications (> to 1200Kbp
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/8/2016