Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20574
Brand Name:BK Virus R-gene
Model:69-013B
Affected:Lot numbers: 1003909940 1004754140 and 1004764070
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:Following a customer compaint about difficulties to interpret low viral load in whole blood samples, an investigation showed that the sensitivity of the BK Virus R-gene® claimed in the IFU for whole blood was not confirmed.
The issue is only impacting whole blood samples with a low viral load, where amplification leads to creeping/flat curves difficult to interpret.
This issue has no impact on negative results or on high positive samples, as the amplification curves are well identified.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/8/2016