Type of Product:Device
Medsafe Reference:20564
Brand Name:Defibrillator, Semi-automatic
Model:
Affected:LIFEPAK CR, Plus AED and LIFEPAK EXPRESS AED
Software version:
Recalling Organisation:Medxus Ltd
Unit 4
100 Fitgerald Avenue
Christchurch 8011
Contact Information:03 365 7403
Manufacturer:Physio-Control Corporation
Issue:The LIFEPAK CR Plus AED or LIFEPAK EXPRESS AED may fail to initiate voice prompts when the ON/OFF button is pressed and the lid is opened due to an internal component (reed switch) that can intermittently become fixed in the closed position.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Public Recall
Date Commenced:25/8/2016