Detail of Recall

Type of Product:Device
Medsafe Reference:20545
Brand Name:Siemens RAPIDPoint 400/405/500 Systems and RAPIDLab 1200 Systemss
Model:
Affected:RAPIDPoint® 400/405/500 Blood Gas Analyzer - RAPIDLab® 1240/1245/1260/1265 Blood Gas Analyzer
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue: In an unlikely set of circumstances, there is a potential for the first and/or last name of one patient to be printed with Patient ID and result data from a different patient, even though those fields have been turned off on the system. The patient ID and patient test result data, however, are correct on the analyser screen and the LIS. Only the printout may contain the incorrect First or Last Name data, which should not have been printed because those fields were turned Off in Setup.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/8/2016