Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20538
Brand Name:UniCel DxH 800 UniCel DxH 600
Model:
Affected:629029, B24465, B24802, B68304 and B23858
Software version:All Software Version
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:There is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID. This issue does not occur if the primary identifier is configured as Specimen ID.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/8/2016