Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20534 |
| Brand Name: | Base for Terumo Advanced Perfusion System |
| Model: | 801764 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Terumo Australia Pty Ltd (NZ) c/- Healthcare Logistics 56 Carrington Road Mount Albert Auckland 1025 |
| Contact Information: | 0800 66 77 57 |
| Manufacturer: | Terumo Cardiovascular Systems Corp |
| Issue: | Manufacturer has received reports of the Central Control Monitor (CCM) for Terumo Advanced Perfusion System 1 losing partial or full functionality. Symptoms that have been associated with a malfunctioning CCM include: full or partial loss of display; loss of control functions; or an inability to distinguish various status, alert, or alarm conditions. In one report, the CCM emitted smoke. Although the occurrence for a CCM malfunction is unlikely, the possible risks of malfunction range from negligible to very severe. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/8/2016 |