Detail of Recall

Type of Product:Device
Medsafe Reference:20522
Brand Name:Elekta Monaco RTP
Model:
Affected:
Software version:3.1 amd higher
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:In a specific workflow where contours are edited (enlarged, moved, copied etc;) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner.
If the patient is not saved and recalled, in a subsequently created plan the BEV will show the contours as still present while the Transverse, Sagittal and Coronal views will show the contours as deleted.
If the user bases their clinical devisions on incorrect contours it could lead to dose erors and potential patient mistreatments.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/8/2016