Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20522 |
| Brand Name: | Elekta Monaco RTP |
| Model: | |
| Affected: | |
| Software version: | 3.1 amd higher |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | In a specific workflow where contours are edited (enlarged, moved, copied etc;) and then the contour is deleted on some slices and saved in the same session, it is possible that the deleted contours are still present which would not be the intent of the planner. If the patient is not saved and recalled, in a subsequently created plan the BEV will show the contours as still present while the Transverse, Sagittal and Coronal views will show the contours as deleted. If the user bases their clinical devisions on incorrect contours it could lead to dose erors and potential patient mistreatments. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/8/2016 |