Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20513 |
| Brand Name: | Merit/47258/Manifold Kits |
| Model: | |
| Affected: | H875600, H895239, H829 |
| Software version: | |
| Recalling Organisation: | InterMed Medical Ltd 71 Apollo Drive Albany AUCKLAND 1311 |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Merit Medical Systems Inc |
| Issue: | Due to supplier moulding defect in the drip chamber prior to use and during prep, when contrast is drawn out of the contrast bottle air may be drawn into the contrast fluid channel from the vent channel resulting in the contrast having excessive air bubbles which can cause the burette to empty quickly and the blue ball to drop to the bottom of the chamber interfering with the contrast administration. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/8/2016 |