Detail of Recall

Type of Product:Device
Medsafe Reference:20505
Brand Name:ResMed Astral Portable Ventilator
Model:
Affected:All Astral 100 and Astral 150 ventilators
Software version:
Recalling Organisation:ResMed (NZ) Ltd
266 Runciman Road
Pukekohe
Auckland
Contact Information:0800 737 635
Manufacturer:ResMed (Australia) Ltd
Brand Name:Resmed Astral Batteries and Battery Packs
Model:
Affected:All Astral External Batteries, All Astral battery packs
Software version:
Recalling Organisation:ResMed (NZ) Ltd
266 Runciman Road
Pukekohe
Auckland
Contact Information:0800 737 635
Manufacturer:ResMed (Australia) Ltd
Issue:There have been a small number of occurrences (<0.1%) where an internal electrical fault has led to ventilation ceasing without either of the low battery alarms being activated. In all of these cases the Total Power Failure alarm is activated as intended. This fault has also been reported with the Astral External Battery, however there have been no reports of cessation of ventilation.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/8/2016