| Type of Product: | Device |
| Medsafe Reference: | 20499 |
| Brand Name: | Terumo Advanced Perfusion System 1 |
| Model: | 801764 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Terumo Australia Pty Ltd (NZ) c/- Healthcare Logistics 56 Carrington Road Mount Albert Auckland 1025 |
| Contact Information: | 0800 66 77 57 |
| Manufacturer: | Terumo Cardiovascular Systems Corp |
| Issue: | Reports have been received of issues that manifest themselves as power or communications failures in the system network. Terumo's investigation revealed an internal cable with an inadequate crimp was installed in a subset of the total population of Terumo System 1 units. The poor crimp in the cable can create a loss or degradation of internal power on one side of the system's base. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/8/2016 |