Detail of Recall

Type of Product:Device
Medsafe Reference:20489
Brand Name:Beckman Coulter UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System
Model:775222
Affected:
Software version:all software revisions
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Manufacturer has received and confirmed a report of a fire within the Stainer module of a customer's DxH Slidemaker Stainer. The fire was fully contained within the Stainer module. The event was preceded by a significant increase in the frequency of the automatic top-off of the module's methanol bath. Beckman Coulter is initiating this corective action to reduce the potential for flammable liquids coming into contact with electrical components in the Stainer module possibly causing a fire.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/8/2016