Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20474 |
| Brand Name: | Welch Allyn- vaginal light source |
| Model: | |
| Affected: | 78800,73410,73412,73414,73416, 78810, 78812,78814, and 78816 |
| Software version: | |
| Recalling Organisation: | Universal Specialities Ltd trading as USL Medical 494 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 09 829 0595 |
| Manufacturer: | Welch Allyn Inc |
| Issue: | KleenSpec Corded Illuminator has the potential to be incorrectly connected to a commonly available AC power cord instead of the illuminator low voltage transformer power cord which is supplied with the illuminator. This may result in a shattered lamp and a potential electric shock. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/8/2016 |