Detail of Recall

Type of Product:Device
Medsafe Reference:20465
Brand Name:Philips IntelliVue Patient Monitors
Model:IntelliVue Patient Monitors
Affected:MX400, MX450, MX500, MX550
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:It has recently been determined that when certain patient monitors (MX400, MX430, MX450, MX500, MX550) have been powered on continuously for at least three months, the waveforms displayed on the monitor may be outdated and therefore fail to reflect the patient's current condition. Should this occur, however, the monitor's alarms and numeric displays of vital signs will continue to function as specified and accurately reflect current patient data.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/8/2016