Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20444 |
| Brand Name: | Roche Accu-Chek Inform 11 Base Unit and Handheld Base Unit |
| Model: | 05060290001 and 04805658001 |
| Affected: | |
| Software version: | 3.00.00 |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | |
| Issue: | Some Accu-Chek® Inform II Base Units (BU) and Handheld Base Units might produce transmission errors in the form of data loss in the USB communication between the meter and the DMS; for example, cobas IT 1000 or any other 3rd party software. This could happen in rare cases when the base unit is connected via USB connection to a PC. This may be indicated by numerous (nonspecific) error messages sent from the meter to the DMS in the form of unexpected software failure 11800 events. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 28/7/2016 |