Detail of Recall

Type of Product:Device
Medsafe Reference:20444
Brand Name:Roche Accu-Chek Inform 11 Base Unit and Handheld Base Unit
Model:05060290001 and 04805658001
Affected:
Software version:3.00.00
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:
Issue:Some Accu-Chek® Inform II Base Units (BU) and Handheld Base Units might produce transmission errors in the form of data loss in the USB communication between the meter and the DMS; for example, cobas IT 1000 or any other 3rd party software. This could happen in rare cases when the base unit is connected via USB connection to a PC. This may be indicated by numerous (nonspecific) error messages sent from the meter to the DMS in the form of unexpected software failure 11800 events.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:28/7/2016