Detail of Recall

Type of Product:Device
Medsafe Reference:20437
Brand Name:Siemens AXIOM Artis, Artis zee and Artis Q/Q Systems
Model:
Affected:multiple
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Brand Name:Siemens AXIOM Artis, Artis zee and Artis Q/Q Systems
Model:
Affected:Multiple
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:Siemens AXIOM Artis, Artis zee and Artis Q/Q Systems
Model:
Affected:Multiple
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens AG
Issue:Contaminants in the form of biomass can develop in the cooling system of Artis systems, which can result in damage to the pump system. The presence of contaminants in the tube cooling circuit impairs the performance of the pump. This fault can even lead to a functional failure of the system. The tube assembly can become overheated so that no radiation can be released.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2016