Type of Product:Device
Medsafe Reference:20436
Brand Name:Maquet Operating Table
Model:
Affected:Fixture (1002.65AO)
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Maquet GmbH
Issue:There has been a report that the welded joint between the square mount and the first fixable swivel joint had broken, caused by ineffective welding in addition to rough handling and multiple overloading.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory