Detail of Recall

Type of Product:Device
Medsafe Reference:20428
Brand Name:Unomedical Frazier Suction Instrument
Model:309.02.006 309.02.008 309.02.008L 309.02.010 309.02.010L 309.02.012 309.02.012L
Affected:All lots
Software version:
Recalling Organisation:Convatec (New Zealand) Ltd
WHK Gosling Chapman
Level 6, 51-53 Shortland Street
Auckland
Contact Information:0800 441 763
Manufacturer:Unomedical Sdn Bhd
Issue:IFU being issued due to complaints from the users in relation to breaking of the instrument at the junction of the handle and the metal tubing when pressure is applied at the tip. If the issue were to occur it would result in a loss of suction and loss of visibility in the surgical field.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/7/2016