| Type of Product: | Device |
| Medsafe Reference: | 20417 |
| Brand Name: | Laerdal Suction Unit |
| Model: | 780800 LSU NiMH |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Laerdal (New Zealand) Ltd 79 Apollo Drive Rosedale Auckland 0632 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Laerdal Medical AS |
| Issue: | There has been a problem identified that is related to use of the Laerdal Suction Unit (LSU) when used with the LSU NiMH Battery that hase been stored or installed at low temperatures for a prolonged period. The problem may occur if the LSU with a NiMH battery, or the NiMH battery, has been stored at low temperatures so that the core temperature of the NiMH battery is low, typically between 0 °C and 9 °C (32 °F to 48 °F). If used with a NiMH battery with low core temperature, the LSU may shut off after some seconds when operated at 350 mmHg or 500+ mmHg (46.6 kPa or 66.5+ kPa) settings. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 3/8/2016 |