Detail of Recall

Type of Product:Device
Medsafe Reference:20411
Brand Name:Medtronic O-arm O2 Imaging System X-ray system
Model:B1-700-02000, B1-700-00027 & B1-700-00028
Affected:
Software version:Software Version 3.1.x, Software Version 3.2
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:The O-arm® system requires updates in order to maintain compliance with applicable performance standards. Specifically,
1.Certain aspects of the technical information in the Instructions for Use document require clarification.
2.Radiation measurement methodologies associated with confirming the accuracy specifications of the air kerma displays do not fully comply with requirements.
3.The O-arm System X-ray technique factor display accuracy related to mA might not be compliant to the system specifications
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Healthcare Professional
Date Commenced:26/7/2016