Detail of Recall

Type of Product:Device
Medsafe Reference:20400
Brand Name:Stryker Exeter Non V40 Rasp Handle
Model:Exeter Non V40 Rasp Handle / 0930-9-003
Affected:All Rasps Handles with Plastic Lever
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:There have been instances where a rivet or rivets have come off an Exeter Non V40 Rasp Handle. These instances are associated with the Exeter Rasp Handles with either the green or black plastic lever, manufactured between January 1995 and October 2005. The design was updated in 2005.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/7/2016