Detail of Recall

Type of Product:Device
Medsafe Reference:20362
Brand Name:Beckman Coulter UniCel DxC Synchrom Systems No Foam Reagent
Model:B64130
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:An iincompatibility between No Foam DxC reagent (REF B64130) and the DxC hydropneumatic parts associated with No Foam delivery has been identified.
DxC No Foam reagent is intended to prevent formation of foam and/or bubbles in the DxC waste system of the hydropneumatic subsystems. Over time, this incompatibility may lead to:
- Breaks in No Foam Cap Assembly
- Cracks in No Foam Waste Collector; and
- Seepage outside No Foam tubing
This issue may lead to cracks in the Waste Collector which may stop Instrument operation due to vacuum error, and potentially lead to biohazardous waste exposure
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/7/2016