Detail of Recall

Type of Product:Device
Medsafe Reference:20357
Brand Name:Sorin 1T /3T FlexTherm Systems
Model:16-02-50 16-02-80 16-02-91 16-02-82 16-02-83 16-02-85 16-02-95 and 16-70-00
Affected:
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Sorin Group Deutschland GmbH
Issue:The Instructions for Use for the 1T and 3T/FlexTherm System prescribe the addition of 50ml and 150ml respectively of medical grade 3% hydrogen peroxide to the filtered tap water in the device, which results in a maximum concentration of approximately 330ppm. The purpose of adding the hydrogen peroxide is to prevent microbial growth between the regular cleaning / disinfection cycles performed every 2 weeks.
The manufacturer has determined that plastic heat-exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow for diffusion of some quantity of hydrogen peroxide during a typical procedure
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/7/2016