Detail of Recall

Type of Product:Device
Medsafe Reference:20295
Brand Name:Beckman Coulter Access Free T3 reagent
Model:A13422
Affected:Lot Nos: multiple (all lots manufactured post June 2015)
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Specific lots manufactured post June 2015 of the Access Free T3 Reagent demonstrate an upward shift in patient results.
Patient sample results will shift upward by approximately 10.4% across the reference interval when compared to the results generated with reagent lots manufactured prior to the June 2015 design change.
Additional assay performance characteristics are not affected.
Due to matrix differences, QC values may not demonstrate a shift.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/6/2016