Detail of Recall

Type of Product:Device
Medsafe Reference:20292
Brand Name:Stryker TRident Constrained Liner Impactor Tips
Model:21992022, 21992028 and 21992032
Affected:All lots affected
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:During steam sterilisation validation the Trident Constrained Liner Impactor Tips do not meet the required Sterility Assurance Level (SAL) of 10-6. These instruments are sterilised at the hospital prior to surgery and are not distributed as sterile.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/6/2016