Type of Product:Device
Medsafe Reference:20289
Brand Name:MOSAIQ with Locate
Model:Mosaiq
Affected:
Software version:2.62
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Impac Medical Systems Inc
Issue:In Treatment Chart window while selecting the Plan-of-the-Day IQ script to run, the treatment plan could be accidently changed by scrolling the mouse wheel up and down and inadvertently selecting another treatment plan.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/6/2016