Type of Product:Device
Medsafe Reference:20286
Brand Name:Siemens Dimension Vista System - Dimension clinical chemistry System
Model:Dimension CRE2, Dimension Vista CRE2
Affected:K1033A, DF33B
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue: Dimension / Dimension Vista Creatinine (CRE2) Assay exhibits a negative bias at the low end of the urine Analytical Measurement Range (AMR). The Limit of Quantitation (LoQ) claim (442 µmol/L) for urine samples is not met. This issue also affects all future lots of CRE2 until a solution is implemented. The serum/plasma CRE2 AMR is not affected by this issue.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/6/2016