Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20281
Brand Name:Draeger Air-Shields Resuscitaire Radiant Warmer
Model:Babytherm 8010
Affected:Heating element
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draeger Medical Systems Inc
Issue:When the Babytherm is operated in skin temperature mode with skin temperature sensor, but without a patient, the automatic controls monitoring skin temperature are not activated and the radiant warmer operates continuously at a high output level. During this forced continuous operation underder high thermal load, the heating element ages faster.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/6/2016