Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20279 |
| Brand Name: | GE MRI system, full-body, superconducting magnet |
| Model: | Discovery MR750w |
| Affected: | MR750w |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w. A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 15/6/2016 |