Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20267 |
| Brand Name: | Sirona Dac Universal Combination Steriliser |
| Model: | |
| Affected: | Serial Nos:14000-99999;104000-199999; 202000-999999; B14000-B99999; B104000-B199999; B202000-B999999 |
| Software version: | |
| Recalling Organisation: | Sirona Dental Systems Ltd Unit 21, 14-22 Tritoin Drive Albany North Shore AUCKLAND |
| Contact Information: | |
| Manufacturer: | Nitram Dental A/S |
| Issue: | Safety concern resulting from Dac Universal MK111 units that have been repaired using spare parts intended for MK11 units. The units possibly at risk are those that: 1. Fall within a determined serial number range and, 2. Where screws for the lid holder originally designed for the MK11 may have been used for a MK111 unit. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/6/2016 |