Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:20263
Brand Name:Beckman Coulter ALT
Model:OSR6007 OSR6107 OSR6507 and OSR6607
Affected:All Lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Brand Name:Beckman Coulter Glucose-STAT
Model:OSR6140
Affected:All lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:Previously unknown drug interferences from Sulfasalazine and Sulfapyridine have been identified.
ALT and Glucose-STAT are IVD products designed for the quantitative determination of Alanine aminotransferase (ALT) or Glucose on the Beckman Coulter AU series of analysers.
There is a possibility that drug interferences from administration of Sulfasalazine to patients may generate false low (ALT) results and false low Glucose-Stat results .
Sulfapyridine may generate false high Glucose-Stat results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/6/2016
Hide menus