Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20263 |
| Brand Name: | Beckman Coulter ALT |
| Model: | OSR6007 OSR6107 OSR6507 and OSR6607 |
| Affected: | All Lots |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Brand Name: | Beckman Coulter Glucose-STAT |
| Model: | OSR6140 |
| Affected: | All lots |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | Previously unknown drug interferences from Sulfasalazine and Sulfapyridine have been identified. ALT and Glucose-STAT are IVD products designed for the quantitative determination of Alanine aminotransferase (ALT) or Glucose on the Beckman Coulter AU series of analysers. There is a possibility that drug interferences from administration of Sulfasalazine to patients may generate false low (ALT) results and false low Glucose-Stat results . Sulfapyridine may generate false high Glucose-Stat results. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/6/2016 |