Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20257 |
| Brand Name: | Stryker Target Nano Detachable Coil |
| Model: | 542102, 542103, 542152, 542153, 542154, 543101, 543102, 543103, 543151, 543152, 543153, 543154, 544203, 544204, 544254, 544304, 544306, 544356, 545202, 545203, 545204, 545206, 545253, 545254, 545256, 545304, 545306, 545308 |
| Affected: | Multiple Lots |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Neurovascular |
| Issue: | Some units of the Target Nano Detachable coils do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacture which can lead to an increased occurrence of coil stretching. Note that this issue relates to the Target Nano range of detachable coils only. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/6/2016 |