Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20250 |
| Brand Name: | ConMed ECG Electrodes |
| Model: | ThermoGard Dial Dispersive Electrode Pads |
| Affected: | CM-51-7710 lots 141010X to 20150705X, CM-7-382 lots 141010X to 201507005X, CM-7-382 lots 141010X 20150705X |
| Software version: | |
| Recalling Organisation: | Paragon Care Group New Zealand Limited 69 Elizabeth Knox Place Glen Innes AUCKLAND |
| Contact Information: | 0508 654 258 |
| Manufacturer: | ConMed Corporation |
| Issue: | ThermoGard Dual Dispersive Electrode Pads are not compatible with some brands of electrosurgical generators using passive CQM (Contact Quality Monitoring) Systems. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/6/2016 |