Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20245 |
| Brand Name: | RaySearch RayStation Treatment Planning System |
| Model: | |
| Affected: | |
| Software version: | 5.0.0.37, 5.0.1.11 and 5.0.2.35 |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | An issue has been found with the treatment plan report in RayStation 5. For a treatment plan with multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose are included as the last dose statistics prior to the plan dose statistics. For such a report template, all reports for multiple beam set plans would always be affected. The user must be aware of information to avoid incorrect assessment of treatment plan dose levels. i.e Inspect all report templates that are intended to be used for plans with multiple beam sets. Make sure that "POI Dose statistics (Plan)" is included immediately before "ROI Dose statistics [Plan dose]" in the Report content area. This issue will be resolved in the next version of RayStation, scheduled for market release Nov 2016 |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/6/2016 |