Detail of Recall

Type of Product:Device
Medsafe Reference:20244
Brand Name:Dimension Vista System - Blood Urea Nitrogen (BUN) Flex
Model:K1021
Affected:16004AB, 16033AC, 16034AA, 16048AA, 16055AA,16055AB, 16062AA,16062AB
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:In February 2016, Siemens Healthcare Diagnostics issued field corrective action letter VC 16-01, informing customers of well-to-well accuracy shifts for specific lots of Dimension Vista BUN causing inaccurate patient and/or Quality Control results. At that time, preliminary investigation demonstrated loss of enzymatic activity of the GLDH reagent wells made from a specific mold/cavity. They have since confirmed complaints on additional Vista BUN Flex cartridge lots, suggesting an additional cause of reduced GLDH activity.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/6/2016