Detail of Recall

Type of Product:Device
Medsafe Reference:20239
Brand Name:TCVS Heart lung machine
Model:Serial interface module RS-232, Serial interface module RS-485, Interface module for CDI System 100/101, Interface module for CDI System 500
Affected:Interface Module
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Terumo Cardiovascular Systems (Terumo CVS) has received reports of malfunctions between 2006 and 2010 involving the loss of data transfer capabilities. Some reports stated that the odor of smoke was associated with the failures. The malfunctions occurred during set-up, during cardiopulmonary bypass, and during routine testing of the module. Terumo CVS' investigation has determined that the malfunctions are caused by a circuit board that failed.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory