Type of Product:Device
Medsafe Reference:20237
Brand Name:TCVS Heart lung machine
Model:802018
Affected:System Flow Module
Software version:
Recalling Organisation:Terumo Australia Pty Ltd (NZ)
c/- Healthcare Logistics
56 Carrington Road
Mount Albert
Auckland 1025
Contact Information:0800 66 77 57
Manufacturer:Terumo Cardiovascular Systems Corp
Issue:Terumo Cardiovascular Systems (CVS) received reports of random flow readings and/or intermittent false back flow alarms for the Terumo Advanced Perfusion System. 1. There was no tubing in the flow sensor. When operating as intended, the module will generate '---' (three dashes) in the display, and issue a 'check sensor' alert message, when there is no tubing in the flow sensor
Recall action type:Correction
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/6/2016