Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 20235 |
| Brand Name: | TCVS Heart lung machine |
| Model: | 801558, 801238 |
| Affected: | Terumo Flexible Halogen Lamp |
| Software version: | |
| Recalling Organisation: | Terumo Australia Pty Ltd (NZ) c/- Healthcare Logistics 56 Carrington Road Mount Albert Auckland 1025 |
| Contact Information: | 0800 66 77 57 |
| Manufacturer: | Terumo Cardiovascular Systems Corp |
| Issue: | A lamp use as an optional accessory with the Terumo® Advanced Perfusion System uses a halogen bulb which will become hot during use. The lamp will become hot, can cause damage or ignition to materials that are placed in close proximity to the hot lamp housing, and can cause burn injury to personnel that come in close contact with the hot lamp housing. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/6/2016 |